4-24 Input and Output Cards
UM4.6:CS6003 Original — December 1991
4.1.7.1 Overview
During the functional test, the calibrator checks each of the four pulse
count input channels. If any channel fails to perform correctly, replace
the pulse count input card. Figure 4-1 shows a pulse count input input
card layout diagram.
X00611–A
NEW STYLE
High Voltage
DS1 Channel 1 LED
DS2 Channel 2 LED
DS3 Channel 3 LED
DS4 Channel 4 LED
DS5 Channel 5 LED
DS6 Channel 6 LED
DS7 Channel 7 LED
DS8 Channel 8 LED
Figure 4-8. Pulse Count Input Card
4.1.7.2 Equipment Required
The following equipment is required:
Type CS6003 Calibrator
Pulse or Waveform Generator
4.1.7.3 Calibration Procedure
Use the following procedure to calibrate the card:
Step 1: Verify that the calibrator ON/OFF switch to OFF.
Step 2: Insert the pulse count input card in the calibrator test slot
labeled MUX.
Step 3: Connect the umbilical cable labeled MUX DI/BREAKOUT
between the connector on the front of the pulse count input
card and the calibrator receptacle labeled BREAKOUT.
Step 4: Turn the calibrator ON.
4