4-13Input and Output Card
Original — December 1991 UM4.6:CS6003
Table 4-7. Test Resistances for RTD Input Cards
Resistance in Ohms by Temperature Range
in
°
F
Temp.
Coefficient
ohms/
°
C
Adjust-
ment
%
Span
–50
to
200
100 to
500
0 to
300
–330
to
140
100 to
600
0 to
1000
Numerical
Display
0.3850 0 82.01 114.68 93.01 18.06 70.95 93.01 0000
ZERO
(1)
0.5 82.29 115.11 93.34 18.61 71.73 94.11 0016
20 93.01 131.74 106.06 40.09 101.74 135.97 0640
40 103.90 148.57 118.97 61.55 131.74 177.48 1280
60 114.68 165.17 131.74 82.55 161.04 217.55 1920
80 125.37 181.55 144.38 103.04 189.65 256.20 2560
SPAN
(2)
100 135.97 197.69 156.90 123.24 217.55 293.43 3200
0.3902 0 81.83 114.88 92.94 16.56 70.61 92.94 0000
ZERO
(1)
0.5 82.11 115.31 93.27 17.13 71.40 94.05 0016
20 92.94 132.17 106.15 39.24 101.76 136.46 0640
40 103.96 149.23 119.22 61.05 132.17 178.52 1280
60 114.88 166.06 132.17 82.28 161.87 219.15 1920
80 125.71 182.65 144.99 103.08 190.86 258.32 2560
SPAN
(2)
100 136.46 199.02 157.67 123.55 219.15 296.05 3200
0.3923 0 81.72 114.96 92.90 16.57 70.44 92.90 0000
ZERO
(1)
0.5 82.00 115.40 93.23 17.14 71.23 94.01 0016
20 92.90 132.34 106.18 39.06 101.77 136.65 0640
40 103.98 149.49 119.33 60.87 132.34 178.95 1280
60 114.96 166.41 132.34 82.17 162.20 219.81 1920
80 125.85 183.10 145.23 103.10 191.35 259.25 2560
SPAN
(2)
100 136.65 199.55 157.98 123.68 219.81 297.24 3200
1. Zero adjustment
2. Span adjustment
Step 10: Set the decade resistor to the 100 percent of span resistance
level listed in Table 4-7. Adjust the SPAN potentiometer for
channel 1 until the calibrator numerical display is 3200.
Step 11: Repeat steps 9 and 10 until no adjustment is required for
either step.
Step 12: Perform a linearity check for channel 1. Set the decade
resistor to the various resistance levels listed in Table 4-7 for
your card, and observe the calibrator numerical display. Verify
that for each resistance, the display is within +
0002 of the
required number listed in the table. If any incorrect numerical
display appears, the RTD input card requires repair.